Emerson 51-DO-03 Oxygen Equipment User Manual


 
MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION
5.6 CHANGING ALARM PARAMETERS
2. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
3. Move the cursor to "Configure" and press Enter (F4).
4. Move the cursor to "Alarms" and press Enter (F4).
5. Nine menu headers relate to alarms. Alarms 1, 2 and 3, each have a
control header and a setup header. Alarm 4 has only a setup header.
The eighth menu header is for configuring the feed limit timer, and the
ninth menu header is for configuring the interval timer.
NOTE
The 54eA analyzer can be ordered with optional TPC control
feature. The version of the 54eA analyzer provided with the
DO-03/04 does not have control. Although the control head-
er appears in this screen, it is not activated.
To access a header, move the cursor to the desired header and press
Enter (F4).
6. Alarm Control Settings:
a. Move the cursor to the desired output control header. Press Enter
(F4).
b. With the cursor on "Activation method" press Enter (F4).
Alarm 1 control
Alarm 1 setup
Alarm 2 control
Exit Enter
Outputs
Alarms
pH
Exit Enter
Output setpoints
Simulated tests
Configure
Exit Enter
1. This section describes how to configure the analyzer alarms. Alarms 1, 2, and 3 can be assigned to dissolved
oxygen or temperature. In addition, alarm 1, 2, or 3 can be configured as a feed limit timer or as an interval
timer (see steps 10 and 11). Alarm 4 is always a fault alarm.
Activation method
Control mode
Exit Enter
31