Emerson 51-DO-03 Oxygen Equipment User Manual


 
MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION
ITEM CHOICES FACTORY SETTINGS
CONFIGURE (continued)
D.Temperature compensation (Section 5.7)
1. Temperature compensation Auto or manual Auto
2. Manual temperature -15 to 130°C 25°C
E. Noise Reduction (section 5.8)
Noise rejection 50 or 60 Hz 60 Hz
F. Barometric Pressure (Section 5.9)
1. Barometric pressure Auto or manual Auto
2. Barometric pressure manual 0.4 - 1.2 atm (or equivalent units) 1.00 atm
3. Barometric pressure units mm Hg, in Hg, bar, kPa mm Hg
G. Main sensor calibration (Section 5.10)
1. Stabilize reading
a. oxygen (ppm) 0 to 20 ppm 0.05 ppm
b. oxygen (% saturation) 0 to 200% 1%
2. Stabilize time 0 - 30 sec 10 sec
3. Sensor zero stabilization value
4. Salinity (parts per thousand) 0.0 - 99.9
o
/
oo 0.0
o
/
oo
H. Security (Section 5.11)
1. Lock all 000-999 (000 disables) 000
2. Lock program 000-999 (000 disables) 000
3. Lock configuration 000-999 (000 disables) 000
TABLE 5-1. Program Settings List (continued)
23