MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION
5.7 TEMPERATURE COMPENSATION AND TEMPERATURE UNITS
1. Refer to Section 6.1 for a discussion of the ways in which temperature affects dissolved oxygen measurement.
pH
Temperature
Noise rejection
Exit Enter
2. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
3. Move the cursor to "Configure" and press Enter (F4).
4. Move the cursor to "Temperature" and press Enter (F4).
5. Use the and keys to move through the list of items. To make a
change press Edit (F4). Use the arrow keys to change settings to the
desired value. Press Save (F4) to store changes.
Auto: In automatic temperature compensation, the analyzer meas-
ures the temperature using an RTD (resistance temperature device)
in the sensor. The analyzer then uses the measured temperature to
calculate the membrane permeability correction and oxygen solubili-
ty factor.
Manual: In manual temperature compensation, the analyzer uses the
temperature entered by the user to calculate the membrane perme-
ability correction and oxygen solubility factor. It does NOT use the
actual process temperature. Do NOT use manual temperature com-
pensation unless the difference between the calibration and meas-
urement temperatures is less than 2°C.
Manual temperature compensation is useful if the sensor RTD
has failed and a replacement sensor is not available.
If Manual temperature compensation is selected, be sure to enter the
desired temperature.