Emerson 51-DO-03 Oxygen Equipment User Manual


 
24
5.1 CHANGING ALARM SETPOINTS
2. Press any key to enter the main menu. Move the cursor to "Program"
and press Enter (F4).
3. Press Enter (F4).
4. Move the cursor to the desired alarm and press Enter (F4).
5. A screen like the one shown will appear. The alarm is a low alarm
and the setpoint is 0.00 ppm. Press Edit (F4). Use the arrow keys to
change the setpoint. Press Save (F4) to store the new value. Press
Exit (F1) to return to the screen in step 4. Choose a new alarm.
Alarm setpoints
Output setpoints
Simulate tests
Exit Enter
Alarm 1 setpoint
Alarm 2 setpoint
Alarm 3 setpoint
Exit Enter
Alarm Low : 0.000 ppm
Exit Edit
1. Before changing alarm setpoints, be sure that alarms are properly configured. See Section 5.6.
MODEL DO-03/04 SECTION 5.0
SOFTWARE CONFIGURATION