76 Field Setup Wizard
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
11. Configure the properties necessary for the AGA8 calculations. This page is only
displayed for Daniel Gas Ultrasonic meters if both temperature and pressure are set to
Live or Fixed and Base condition correction is selected on the Startup Page.
Configuration parameters include:
Calculations performed internally (by the meter) or Externally
AGA8 method - Gross Method 1, Gross method 2 or Detail
Base temperature and pressure
Specific gravity - reference temperature and pressure
Volumetric gross heating value and reference temperature
Molar density reference temperature and pressure
Flow Mass density, flow compressibility and Base compressibility
Gas composition inputs - components and mole percent
Click Next to continue to the Continuous Flow Analysis page, if View Continuous Flow
Analysis setup was selected on the Startup page.
12. Configure the flow analysis parameters shown below.
Set low and high flow limits
Enable SOS comparison (requires AGA 8 Detail method)
Enable liquid detection and Profile factor limit
Enable Blockage. Enter the percent for Symmetry, Cross-flow, Chords A to D
turbulence
Enable Internal bore buildup
Click Next to continue to the Local display page, if View local display setup was selected
on the Startup page.
13. Configure the parameters for the local display. Use the drop-down arrow in the Display
Items list box and select or modify the parameters that will be displayed; the Display
items, the Display units and the Scroll delay.