96 Using a Field Communicator to configure the meter
Section 4: Configuration Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual
June 2014 3-9000-759 Rev D
Procedure
1. Remove electrical power to the meter. If installed, remove the endcap security latches
and seals and then, remove the endcap.
2. Refer to the Field Communicator Users Manual wiring diagrams and commissioning
instructions provided with your handheld device. Register the product to activate the
end user license.
3. Fully charge the Field Communicator battery prior to use. Important: Do not change
the battery in a hazardous area environment. The power supply is not intrinsically safe
4. On the meter, run the wires through the field wiring conduit and into the transmitter
electronics enclosure.
Figure 4-19 3414 transmitter field wiring conduit entries
A. Field wiring conduit entries (4)
A.