Daniel 3410 Series Ultrasonic Gas Flow Meter Installation Manual Section 4: Configuration
3-9000-759 Rev D June 2014
Daniel MeterLink Setup 73
Section 4: Configuration
After the mechanical and electrical installation is complete and connectivity is established, use
the Daniel MeterLink
Software for Gas and Liquid Ultrasonic Meters Quick Start Manual
(P/N 3-9000-763) to setup software communications with the meter.
4.1 Daniel MeterLink Setup
1. Review the software operating system, hardware and peripheral requirements.
2. Follow the installation instructions for your operating system (Windows® XP, Windows
Vista®, Windows® 7 or Windows® 8).
3. Configure a direct connection driver for first time modem configuration for Daniel
MeterLink communications.
4. Select the Daniel MeterLink
desktop icon and complete the information in the
Registration Wizard to keep you informed about updates and technical support.
Registration options include:
Register by e-mail
Register by phone
Register later (remind me)
Register later using the Help|Resister Program menu (don’t remind me)
5. Select
File|Program Settings and customize the user-preferences (e.g. User name,
Company name, display units, Liquid Meter volume units and other interface settings)
6. Connect to your meter. If your meter is not shown in the list, select
Edit Meter Directory
and setup the connections properties.
7. Save the meter configuration file, collect a Maintenance log and Waveforms to
document the meter’s “As Found” settings.
8. Run the Field Setup Wizard.