Chapter 3. Programming Site Data
24 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
3.2 Entering Transducer Parameters (cont.)
Note: The choices made early in the Transducer and Pipe options
determine the prompts available later. If the PT878 does not scroll to
a particular parameter, it is not necessary for that transducer or pipe
type. For example, the Lining window is not available if you select a
wetted transducer.
2. The next prompt asks you to enter the transducer number (printed on
the transducer itself), or to specify that you are using a special
application transducer.
a. From the Type prompt, press the [T] key to reach the Transducer
prompt, and press
[ENTER].
b. A drop-down list of transducer numbers opens as in Figure 11 on
page 25. (The list varies, depending on whether you have selected
wetted or clamp-on in the previous prompt. See Table 4 on page 25.)
Press the [
T] or [S] keys to scroll to the appropriate number, or
scroll to “Special” for a special application transducer. To speed
scrolling, you can press the [
X] key to scroll down by a page, or the
[
W] key to scroll up by a page.
c. Press [ENTER] to confirm your selection.
The program now varies, depending on whether you have selected standard
or special transducers.
If you have selected a standard wetted or clamp-on transducer, the
PT878 comes programmed with the needed parameters. Proceed to
Confirming Entries on page 30.
However, if you have selected a special application transducer, go to
page 26.