GE PT878 Oxygen Equipment User Manual


 
Chapter 8. Servicing the PT878
190 Transport® Model PT878 Portable Liquid Flowmeter User’s Manual
8.10.3 Testing the Watchdog Timer Circuit
The PT878 includes a watchdog timer circuit. If a software error causes the
meter to stop responding, this circuit automatically resets the meter. A
properly functioning PT878 restarts if you run the Watchdog Test.
To start the Watchdog Test, scroll to the Watchdog Test option in the Test
Menu and press
[ENTER]. The screen appears similar to Figure 126 below.
Figure 126: Watchdog Test Screen
Press
[F2] (No) to cancel the test and return to the Menu screen, or press [F3]
(Yes) to start the test. The PT878 should go blank for a few seconds, and
then restart. If it does not follow this sequence, consult the factory.