Emerson 5081FG Oxygen Equipment User Manual


 
Instruction Manual
IB-106-5081, Rev. 1.0
May 2005
Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-13
Model 5081FG
%
CELL
TEMPERATURE
MAXIMUM
26020021
Figure 3-26. MAX CELL T
g. MAX CELL T
Refer to Figure 3-26. Use this screen to ex-
amine the maximum temperature attained
by the O
2
cell. This value can be reset under
the PROGRAM MENU. Pressing NEXT
accesses the SW VER screen.
h. SW VER
Use this screen to see the software version
number for the Model 5081 Transmitter.
Pressing NEXT accesses the UNIT SER #
screen.
i. UNIT SER #
Use this screen to see the unit serial num-
ber for the Model 5081 Transmitter. Press-
ing NEXT accesses the SW BUILD
NUMBER screen.
j. SW BUILD NUMBER
Use this screen to see the software build
number for the Model 5081 Transmitter.
Pressing NEXT accesses the SW BUILD
DATE screen.
k. SW BUILD DATE
Use this screen to see the software build
date for the Model 5081 Transmitter.
Pressing NEXT returns to the beginning of
the DIAGNOSTICS MENU branch (the
SHOW FAULT screen).
3-7 CALCHECK MENU
The CALCHECK MENU branch of the menu
tree (Figure 3-5) allows you to perform a cali-
bration check of the analyzer. Before performing
a calibration check, ensure the high calibration
check gas and low calibration check gas O
2
percentages are entered into the electronics via
the PROGRAM MENU. To set these values,
refer to paragraphs 3-5j and 3-5k.
Once these values are set, access the CAL-
CHECK MENU branch by pressing CAL on the
IRC when in the PROCESS DISPLAY screen
(Normal or Faulted). Each screen in this branch
identifies a process step in the calibration check
procedure. The first screen in the sequence is
the IN MANUAL? screen.