Emerson 5081FG Oxygen Equipment User Manual


 
Instruction Manual
IB-106-5081, Rev. 1.0
May 2005
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 5081FG
The Model 5081 Transmitter is a two-wire
transmitter providing an isolated output, 4-
20 mA, that is proportional to the measured
oxygen concentration. A customer-supplied
24 VDC power source is required to simul-
taneously provide power to the electronics
and a 4-20 mA signal loop. The transmitter
accepts millivolt signals generated by the
probe and produces the outputs to be used
by other remotely connected devices. The
output is an isolated 4-20 mA linearized
current.
d. System Features
1. The cell output voltage and sensitivity
increase as the oxygen concentration
decreases.
2. High process temperatures eliminate
the need for external cell heating and
increase cell accuracy.
3. HART communication is standard. To
use the HART capability, you must
have either:
(a) HART Model 275/375
Communicator
(b) Asset Management Solutions
(AMS) software for the PC
4. Easy probe replacement due to the
light-weight, compact probe design.
5. Remote location of the Model 5081
Transmitter removes the electronics
from high temperature or corrosive
environments.
6. Power is supplied to the electronics
through the 4-20 mA line for intrinsic
safety (IS) purposes.
7. Infrared remote control (IRC) allows
interfacing without exposing the
electronics.
8. An operator can operate and diagnos-
tically troubleshoot the Two-Wire In
Situ Oxygen Analyzer in one of two
ways:
(a) Infrared Remote Control. The IRC
allows access to fault indication
menus on the Model 5081 Trans-
mitter LCD display. Calibration can
be performed from the IRC keypad.
(b) Optional HART Interface. The
Two-Wire In Situ Oxygen Ana-
lyzer’s 4-20 mA output line trans-
mits an analog signal proportional
to the oxygen level. The HART
output is superimposed on the 4-20
mA output line. This information
can be accessed through the
following:
1 Rosemount Analytical Model
275/375 Handheld Communi-
cator - The handheld commu-
nicator requires Device
Description (DD) software
specific to the Two-Wire In
Situ Oxygen Analyzer. The
DD software will be supplied
with many Model 275/375
units but can also be pro-
grammed into existing units at
most Fisher-Rosemount Ana-
lytical service offices. See
Section 4, HART/AMS, for
additional HART
information.
2 Personal Computer (PC) -
The use of a personal com-
puter requires AMS software
available from Fisher-
Rosemount.
9. Selected Distributed Control Systems -
The use of distributed control systems
requires input/output (I/O) hardware
and AMS Security codes are provided
to (by infrared remote control) prevent
unintended changes to analyzers adja-
cent to the one being accessed.
10. A calibration check procedure is provided
to determine if the Rosemount Analytical
Two-Wire In Situ Oxygen Analyzer is cor-
rectly measuring the net oxygen concen-
tration in the industrial process.