Leica M205 FA Microscope & Magnifier User Manual


 
Leica M-series User Manual 52
Under certain circumstances, the light source
must not be moved along with the optics
carrier. A typical example is multifocus images
in which the Z-stack changes while the angle
of incidence of the light must remain the same.
For such purposes, the Leica LED5000 MCI™ is
directly fastened to the column.
Installation on the column
1. Pull the retaining stirrup out of the Leica
LED5000 MCI™.
2. Screw the retaining stirrup into the column
at the notch using the single screw. If you
tighten the screw, the retaining stirrup
is automatically moved into the correct
position.
3. Push the Leica LED5000 MCI™ onto the
retaining stirrup.
Leica LED5000 MCI™: Alternative Assembly