Leica EL 6000 – Troubleshooting
Fault description Possible faults Remedy
Power indicator is not illuminated after being
switched on, instrument shows no function
No power supply voltage present; connecting
cable not plugged in correctly/defective
Check power supply voltage and connecting
cable
Fuses in instrument defective Replace fuses, use permitted replacement
type only (see Page 113)
Power switch is illuminated after being switched
on, instrument shows no function
Housing cover loose
or removed
Attach housing cover securely
Lamp worn out Replace lamp
Target system receives too little light or none at all
Shutter closed or intensity switch at left limit
stop
Open shutter or rotate intensity switch
towards the right
Light guide not inserted into light guide
output properly
Check light guide
Light guide damaged/defective Repair or replace
If these actions do not fix the problem, send the Leica EL6000 compact
light source to Leica service with an exact description of the malfunction.