Leica M205 FA Microscope & Magnifier User Manual


 
Leica M-series User Manual 22
Depending on the work situation, it is
useful to restrict the maximum travel
path of the stereomicroscope. This prevents
injury during handling of the sample caused
by fingers or hands becoming trapped or the
specimen touching the objective or even being
damaged by it.
Readjusting the motorized focus
The motorized focus is factory-adjusted and
normally does not need to be readjusted—
even if the maximum travel path is changed.
•
Exception: If the power fails while the
motorized focus is moving, the position
data are lost. In this case, the calibration must
be repeated using the Leica LAS software or the
Leica SmartTouch™. To do so, please consult the
respective manual.
Restricting the bottom travel range
1. Move the motorized focus into the lowest
position you want to reach.
2. Unscrew the screw of the limit stop on the
side of the focusing column.
3. Push the limit stop to the height of the
motorized focus.
It is easiest to move the limit stop by keeping the
screwdriver inserted and moving it upwards.
4. Tighten the screw of the limit stop.
Motorized Focus: Restricting the Travel Path