Cleaning and Maintenance
•
Before carrying out cleaning tasks,
switch off the Leica EL6000 compact
light source.
Use only a dry, lint-free cloth for cleaning. When
cleaning, ensure that no dust or other foreign
bodies enter the device through the ventilation
slots.
In case of damage (even external damage),
immediately stop using the Leica EL6000
compact light source and send it to service.
If it will not be used for a long time, protect the
Leica EL6000 compact light source from dust
using a suitable cover (such as the plastic bag
from the original packaging).
Spare parts and accessories
The only parts of the Leica EL6000 compact
light source that may be replaced by the user
are the instrument fuses and the lamp.
The replacement of the lamp is described
on Page 61.
Replacing the instrument fuse
1. Disconnect the Leica EL6000 from the
power supply.
2. Open the drawer with the instrument fuses
located close to the power supply input.
3. Replace the instrument fuse.
Repairs
Opening or working on the Leica EL6000
compact light source in any manner other than
described above (changing lamps, changing
fuses) may be carried out by the supplier of the
instrument or the supplier's authorized repre-
sentatives.
Permitted replacement fuses
Fuse: 5×20, 2.5 A, slow-blowing, breaking capacity H
Wickmann Series 181 www.wickmann.com
Schurter Series SPT 5×20 www.schurter.ch