The Leica LED5000 MCI™ (for "Multi Contrast
Illumination") is installed using two screws. For
optimum accessibility, the optics carrier should
be removed during assembly (see Page 23.)
Constraints
The Leica LED5000 MCI™ cannot be used
together with the objective nosepiece.
Assembly
•
1. Hold the LED5000 MCI™ with one
hand and tightly screw the retain-
ing stirrups on both top holes on the drive
housing.
2. Connect the CAN-bus cable provided to
either of the two sockets. (The flat part of
the plug must be facing downwards.)
3. Plug the other end of the cable into one of
the two "CTL2" sockets on the column.
Leica LED5000 MCI™