Ensure that the instruments are unpacked on a
level, adequately sized, and nonskid underlay.
Stage assembly
The Leica TL RC™/ RCI™ transmitted-light base
can be equipped with three different Leica
stages. The selected stage is mounted on the
base before commissioning. You can switch
between the stages at any time with just a few
hand movements.
The following paragraph assumes use of the
base without the stage mounted. Disassembly
is performed in reverse order of the following
steps.
Standard stage
1. Take the glass plate from the rectangular
gap in the standard stage.
2. Position the stage on the transmitted-light
base in such way that the four holes align
over those in the base.
3. Attach the stage to the base with the four
supplied hexagon-head screws.
4. Insert the glass plate back into the standard
stage.