Leica M205 FA Microscope & Magnifier User Manual


 
Leica M-series User Manual 144
Leica EL6000 – Specifications (cont'd.)
Light output
Design Type: Storz, long
Special feature Self-locking if no light guide is
plugged in
Fan
Switch-on delay Approx. 60 s after lighting the
lamp
Interface
Type Remote control input for
internal shutter
Design 9-pin D-Sub socket
Remote signal level
Shutter closed –8 to +0.8 V
Shutter open +2.4 to +8 V
Input current Approx. 1 mA
Footswitch
Shutter closed Connection between
pin 2 and 4
Current Approx. 1.2 mA
Shutter frequency Max. 10 Hz
Protective functions
Safety circuit If the cover is open, all components in the instru-
ment (except for the power input and power plug)
are disconnected from the power supply voltage.
Overheating
protection
In the event of overheating, the lamp supply is shut
off. After cooling off, the lamp power supply in the
Leica EL6000 compact light source is automatically
switched on again.
Protection screen If no light guide is plugged into the light output of
the instrument, the light output is automatically
closed off so that no light can escape.