Invacare 1150704 Personal Lift User Manual


 
SECTION 8—ACCESSORIES
Part No 1150704 31 Jasmine™ Patient Lift
Calibrating the Jasmine Scale
NOTE:TheJasmineScalewillbeprecalibratedatthefactorywiththeloadcell.Shoulditbenecessarytorecalibratethescale,fol low
theinstructionsoutlinedbelow.
1. ThepatientandtheslingMUSTberemovedfromthescaletoproperlycalibratetheJasmineScale.Forremovingthe
patientinstructions
refertoOperationonpage 12.
ForslingdetachmentinstructionsrefertoLiftingthePatient
onpage 14.
2. WiththeJasmineScaleon,removethefourscrewsonthebackoftheenclosureandremovethefrontcovertoexpose
thePCboard.
3. PresstheCALbuttonlocatedonthePCboard.TheCALswitchisnotlabeledbutistheonlybuttononthePCboard
locatedinthelowerrightcorner.OncetheCALbuttonispushedʺCAL1ʺwillbeseeninthedisplaywindow.
NOTE:CAL1selectsthecalibrationmodeusing50poundsofcalibratedweight.CAL2selectsthecalibrationmodeforusewith200
poundsofcalibratedweight.PressingtheUNITSkeytogglesbetween
CAL1andCAL2.
4. Whenthedesiredcalibrationmodeisdisplayed,presstheLOCK/UNLOCKkey.
NOTE:ThedisplaywindowwillnowshowUnLd.
5. EnsurethatthereisnoloadonthescaleandpresstheZEROkey.
6. Thedisplaywindowwillshowadashedline(‐‐‐‐‐‐)scrollingacrossandthentheword
LOADwillappearinthedisplay
window.
7. Performone(1)ofthefollowing:
A. FORCAL1OPTION‐support50poundsofcalibratedweightfromthescaleandpresstheZEROkey.
B. FORCAL2OPTION‐support200poundsofcalibratedweightfromthescaleandpresstheZEROkey.
8. Thedisplaywindowwillshowa
dashedline (‐‐‐‐‐‐)scrollingacrossandthenthewordDONEwillappearinthedisplay
window.
9. Removetheweightfromthescale andpresstheLOCK/UNLOCKkey.
NOTE:Theunitwillnowbeinatemporarytestmodeandwillnotlocktheweightdisplay.Thiswillallowweightto
beloadedand
unloadedtocheckthecalibration.
10. TurntheunitoffbypressingtheON/OFFkey.
11. TurntheunitonbypressingtheON/OFFkey.
NOTE:Scalewillnowbeinnormaloperation.
Troubleshooting
Display Codes
CALIBRATIONREQUIRED‐Indicatesimproperstoredcalibrationdata,calibrationisnecessary.
OVERCAPACITY‐Indicatesaweightexceedingthecapacityhasbeenloadedonthescale.
SYMPTOM PROBABLE CAUSE SOLUTION
Unit does NOT work properly. Battery failure. Check battery. Replace if necessary.
Battery has been replaced and unit still
does NOT work properly.
Contact Invacare for Service at
1-800-333-6900
CAL
OCAP