Invacare 1150704 Personal Lift User Manual


 
SECTION 7—MAINTENANCE
Jasmine™ Patient Lift 26 Part No 1150704
Checking and Tightening Mast Pivot
NOTE:Forthisprocedure,refertoFIGURE 7.5.
1. Togainaccesstotheattachinghardware,removethecoversthatareoneithersideofthemast.
2. Checkthatthelocknutistightandsecureonthescrew.
3. Ifneeded,dooneormoreofthefollowing:
•Tightenlocknutandbackoffthelocknut1/8
ofaturn.
•Replacethelocknut.
FIGURE 7.5 Checking and Tightening Mast Pivot
Boom
Washer
Cover
Mast Connector
Locknut
Bushing
Washer
Screw
Cover
NOTE:Locknutsthissideofmastnotshown.
Mast
Screws
Nylon Washer
Teflon Bearing
Teflon Bearing
Nylon Washer