Invacare 1150704 Personal Lift User Manual


 
SECTION 6—TROUBLESHOOTING
Jasmine™ Patient Lift 20 Part No 1150704
SECTION 6—TROUBLESHOOTING
NOTE:Ifproblemsarenotremediedbythesuggestedmeans,pleasecontactyourdealerorInvacare.
SYMPTOMS FAULTS SOLUTION
Patient Lift feels loose. Mast/Base joint loose.
Tie - Rods are loose.
Refer to Replace the locknut. on page 26.
Noisy or dry sound from pivots. Needs lubrication. Refer to Lubricating the Lift
on page 22.
Electric actuator fails to lift when button
is pressed.
Hand-control or actuator connector
loose.
Battery low.
RED emergency stop button pressed IN.
Battery not connected properly to con-
trol box.
The connecting terminals are damaged.
Electric actuator in need of service or
load is too high.
Charge batteries. Refer to Charging the
Battery on page 13.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the control box.
Refer to Charging the Battery
on page 13.
Replace the battery pack. Refer to Charg-
ing the Battery on page 13.
Refer to Replacing the Boom Actuator
on
page 22. Contact your Dealer.
Unusual noise from actuator. Actuator is worn or damaged or spindle
is bent.
Refer to Replacing the Boom Actuator
on
page 22.
Lift arms will not lower in uppermost
position.
Lift arms require a minimum weight load
to lower from the uppermost position.
Pull down slightly on the lift arms.
Lift arms will not lower during a power
retraction.
Shoulder bolt at the junction of the boom
and mast may not be properly installed.
Refer to Replacing the Swivel Bar or
Optional Cradle Attachment on page 25.