AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 87 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare 14 June, 2007
6.1.1.4.2 MG Modules
Table 6.1-7: MG Module of Created SOP Instances
Attribute Name Tag VR Value
Presence
of Value
Source
Series
Mammography Series
Modality (0008, 0060) CS MG ALWAYS Config
Request Attributes Sequence (0040, 0275) DS
Sequence that contains
attributes from the
Imaging Service Request
ANAP MWL
Image
Mammography Image
Positioner Type (0018, 1508) CS
Mammographic
None Always Auto
Distance Source to detector (0018, 1110) DS ANAP Auto
Distance Source to patient (0018, 1111) DS ANAP Auto
Positioner Primary Angle (0018, 1510) DS ANAP User/XRDI
14
Positioner Secondary Angle (0018, 1511) DS ANAP User/XRDI
26
Image Laterality (0020, 0062) CS
R = right
L = left
B = both (e.g. cleavage)
ALWAYS Auto
Organ Exposed (0040, 0318) CS BREAST ALWAYS Auto
Implant Present (0028, 1300) DS
YES
NO ANAP Config
View Code Sequence (0054, 0220) CS Always Config / User
View Modifier Code Sequence (0054, 0222) SQ ANAP Config / User
6.1.1.5 GSPS
6.1.1.5.1 GSPS IOD
Table 6.1-8: IOD of Created DX Image SOP Instances
IE Module Reference Presence of Module
Patient Patient Identification
Table 6.1-1
ALWAYS
Extended Patient
Table 6.1-1
If received from MWL
Study General Study
Table 6.1-1
ALWAYS
Patient Study
Table 6.1-1
ALWAYS
Series General Series
Table 6.1-1
ALWAYS
Presentation Series Table 6.1-9 ALWAYS
Equipment General Equipment
Table 6.1-1
ALWAYS
Image Presentation State
Table 6.1-9
ALWAYS
Display Shutter
Table 6.1-9
When shutter is applied
Displayed Area
Table 6.1-9
ALWAYS
Graphic Annotation
Table 6.1-9
When graphic annotations are present
Graphic Layer
Table 6.1-9
When graphic annotations are present
14
Value is supplied by user of XRDI component in case of XRDI configuration