AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 45 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare 14 June, 2007
2.2.3 RIS Application Entity Specification
2.2.3.1 SOP Classes Supported
This Application Entity provides Standard Conformance to the following SOP Class(es):
Table 2.2-37: SOP Class(es) for the DICOM Store Application Entity
SOP Class Name SOP Class UID SCU SCP
Modality Worklist Information Model – FIND 1.2.840.10008.5.1.4.31 Yes No
2.2.3.2 Association Establishment Policies
2.2.3.2.1 General
The DICOM standard Application context is always proposed:
Table 2.2-38: DICOM Application Context
Application Context Name 1.2.840.10008.3.1.1.1
2.2.3.2.2 Number of Associations
"Workstation" initiates one association at a time to query the worklist.
Table 2.2-39: Number of Associations as an Association Initiator for RIS AE
Maximum number of simultaneous associations initiated
1
2.2.3.2.3 Asynchronous Nature
Table 2.2-40: Asynchronous Nature as an Association Initiator for RIS AE
Maximum number of outstanding asynchronous transactions
1
"Workstation" does not support asynchronous communication (multiple outstanding transactions over a
single connection)
2.2.3.2.4 Implementation Identifying Information
Table 2.2-41: DICOM implementation Class and Version for DICOM RIS AE
Implementation Class UID
1.3.51.0.1.3
Implementation Version Name
AGFA DTF1.0.XX
4
4
XX is the build version number