2.2.2 DICOM Print Application Entity Specification
2.2.2.1 SOP Classes Supported
This Application Entity provides Standard Conformance to the following SOP Class(es):
Table 2.2-18: SOP Class(es) for the DICOM Print Application Entity
SOP Class SOP Class UID SCU SCP
Verification 1.2.840.10008.1.1 Yes No
Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Yes No
Basic Film Session SOP Class 1.2.840.10008.5.1.1.1 Yes No
Basic Film Box SOP Class 1.2.840.10008.5.1.1.2 Yes No
Basic Grayscale Image Box SOP Class 1.2.840.10008.5.1.1.4 Yes No
Printer SOP Class 1.2.840.10008.5.1.1.16 Yes No
Print Job SOP Class 1.2.840.10008.5.1.1.14 Yes No
Presentation LUT SOP Class 1.2.840.10008.5.1.1.23 Yes No
2.2.2.2 Association Establishment Policies
2.2.2.2.1 General
The DICOM standard Application context is always proposed:
Table 2.2-19: DICOM Application Context
Application Context Name 1.2.840.10008.3.1.1.1
2.2.2.2.2 Number of Associations
"Workstation" initiates one association at a time for each destination to which a print request is being
processed in the active job queue list. Only one job per destination will be active at a time, the other remains
pending until the active job for that destination is completed or failed. There can however be several
simultaneous associations to different destinations.
Table 2.2-20: Number of Associations as an Association Initiator for DICOM Print AE
Maximum number of simultaneous associations initiated
1 per destination
(32 max)
2.2.2.2.3 Asynchronous Nature
Table 2.2-21: Asynchronous Nature as an Association Initiator for DICOM Print AE
Maximum number of outstanding asynchronous transactions
None
DICOM Print AE allows a single outstanding operation on any association. Therefore it does not support
asynchronous operations window negotiation, other than the default as specified by the DICOM specification.