AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 66 of 92
Document No. 001132, Revision 2.0
Agfa HealthCare NodeID Livelink : 14140190 14 June, 2007
Parameter Configurable
(yes/no)
Default value
Timeout waiting for a Storage Commitment
Notification (maximum duration of applicability for a
Storage Commitment Transaction UID).
Yes
Modality Worklist Parameters
Modality Worklist SCU time-out waiting for the final
response to a C-FIND-RQ
No 15 minutes
Maximum number of Worklist Items
No 400
Supported Transfer Syntaxes for Modality Worklist
No
Implicit VR Little Endian
Explicit VR Little Endian
Delay between automatic Worklist Updates
Yes
MPPS Parameters
Enable MPPS Reporting
Yes
MPPS SCU time-out waiting for a response to a N-
CREATE-RQ
No 15 minutes
MPPS SCU time-out waiting for a response to a N-
SET-RQ
No 15 minutes
Supported Transfer Syntaxes for MPPS
No
Implicit VR Little Endian
Explicit VR Little Endian
Print Parameters
Print SCU time-out waiting for a response to a N-
CREATE-RQ
No 10 minutes
Print SCU time-out waiting for a response to a N-
SET-RQ
No 10 minutes
Print SCU time-out waiting for a response to a N-
ACTION-RQ
No 10 minutes
Supported Transfer Syntaxes (separately
configurable for each remote printer)
No
Taken from printer device
model provided by Agfa.
Number of times a failed print-job may be retried No Automatically : 8 times
Delay between retrying failed print-jobs No
2 times immediately, then
the job gets parked for 3
minutes and retried 3
times again. In case the
job still fails, it is parked
for 5 minutes and retried 3
times.
Print SCU time-out waiting for a response to a N-
CREATE-RQ
No 10 minutes