AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 69 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare 14 June, 2007
Exceptions exist for attributes which are necessary to correctly import or interpret the image. Therefore the
following values are kept or replaced by generated values:
Table 3.2-2: Necessary Values when performing an Anonymous Export.
Attribute Tag Set to …
Patient’s Name (0010,0010) “Anonymous”
Study Instance UID (0020,000D) Keep value
Modality (0008,0060) Keep value
Series Instance UID (0020,000E) Keep value
Laterality (0020,0060) Keep value
Series description (0008,103E) Keep value
Protocol name (0018,1030) Keep value
Body Part Examined (0018,0015) Keep value
Smallest Pixel Value in Series (0028,0108) Keep value
Largest Pixel Value in Series (0028,0109) Keep value
Presentation Intent Type (0008,0068) Keep value
SOP Instance UID (0008,0018) Keep value
SOP Class UID (0008,0016) Keep value
Specific character set (0008,0005) Keep value
Patient Orientation (0020,0020) Keep value
Image type (0008,0008) Keep value
Image comments (0020,4000) Keep value
Presentation LUT shape (2050,0020) Keep value
3.2.1.1.1.1 Media Storage Application Profile
The Offline-Media Application Entity supports the STD-GEN-CD Application Profile.
3.2.1.1.1.1.1 Options
The Offline-Media Application Entity supports the SOP Classes and Transfer Syntaxes listed in the Table
below:
Table 3.2-3: IOD’S, SOP Classes and Transfer syntaxes for offline media
IOD SOP Class UID Transfer Syntax Transfer Syntax UID
CR Image Storage 1.2.840.10008.5.1.4.1.1.1
Explicit VR Little Endian 1.2.840.10008.1.2.1
Digital X-Ray Image
Storage – For Presentation
1.2.840.10008.5.1.4.1.1.1.1
Explicit VR Little Endian
1.2.840.10008.1.2.1