2.2.2.3.1.2 Proposed Presentation Contexts
The DICOM Print AE is capable of proposing the Presentation Contexts shown in the following table:
Table 2.2-23: Presentation Contexts Proposed by DICOM Print AE
SOP Class Transfer Syntax
Extended
Negotiation
SOP Class SOP Class UID Name UID
Role
Verification 1.2.840.10008.1.1 DICOM Implicit
VR Little Endian
1.2.840.10008.1.2 SCU None
Basic Grayscale Print Management
Meta SOP Class
1.2.840.10008.5.1
.1.9
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2 SCU None
Print Job SOP Class 1.2.840.10008.5.1
.1.14
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2 SCU None
Presentation LUT SOP Class 1.2.840.10008.5.1
.1.23
DICOM Implicit
VR Little Endian
1.2.840.10008.1.2 SCU None
Note:
The Presentation Context shall use Abstract Syntax IDs which correspond to the SOP Classes UID of
the Meta SOP Class specified in the first column of the Transfer Syntax Table or included SOP
Classes. None of the included SOP Classes supports extended negotiation.
2.2.2.3.1.3 SOP Specific Conformance - Basic Grayscale Print Management
Meta SOP Class
"Workstation" provides Standard conformance to the DICOM Basic Grayscale Print Management Class
(1.2.840.10008.1.1.9) as SCU.
Support for Basic Grayscale Print Management as SCU also implies support for the following SOP Classes
as SCU:
1. Basic Film Session SOP Class
2. Basic Film Box SOP Class
3. Basic Grayscale Image Box SOP Class
4. Printer SOP Class
2.2.2.3.1.3.1 Basic Film Session SOP Class (1.2.840.10008.5.1.1.1)
The Basic Film Session IOD describes the presentation parameters which are common for all the films of a
film session (e.g. number of films, film destination, …)
The Basic Film Session SOP Instance refers to one or more Basic Film Box SOP Instance (§2.2.2.3.1.3.2).
"Workstation"(SCU) can send the following DIMSE services: