AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 61 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare 14 June, 2007
PI M-Lut Rescale VOI-Lut P-LUT MU1 MU2
(1) Intercept Slope Type Explanation WinCenter WinWidth
LUT-
Shape (2) (3)
M2 3000 -0.684 OD - 1600 2800 - X X
M1 0.0 1.0 P-VALUES - 2048 4096 INVERSE X X
12-bit P-Value
M2 0.0 1.0 P-VALUES - 2048 4096 IDENTITY X X
M1 0.0 1.0 OD REL - 2048 4096 - X X
12-bit OD REL
M2 0.0 1.0 OD REL - 2048 4096 - X X
M1 0.0 1.0 LOG_E REL E25, NK5. … - (7) - (7) - X
12-bit Image --> VOI Lut OD REL
(12)
M2 0.0 1.0 LOG_E REL E25, NK5. … - (7) - (7) - X
M1 0.0 1.0 LOG_E REL E25, NK5. … - (7) - (7) - X
12-bit image --> VOI Lut
ODGamma (12)
M2 0.0 1.0 LOG_E REL E25, NK5. … - (7) - (7) - X
M1 0.0 1.0 P-VALUES E25, NK5. … - (7) - (7) INVERSE X
15-bit Image --> VOI Lut P-Value
(15)
M2 0.0 1.0 P-VALUES E25, NK5. … - (7) - (7) IDENTITY X
M1 0.0 1.0 LOG_E REL - - (7) - (7) - X
12-bit Image --> VOI Lut OD REL
(12)
M2 0.0 1.0 LOG_E REL - - (7) - (7) - X
M1 0.0 1.0 LOG_E REL - - (7) - (7) - X
12-bit Image --> VOI Lut
ODGamma (12)
M2 0.0 1.0 LOG_E REL - - (7) - (7) - X
M1 0.0 1.0 P-VALUES - WC (9) WW (9) INVERSE X
15-bit Image --> VOI Lut P-Value
(15)
M2 0.0 1.0 P-VALUES - WC (9) WW (9) IDENTITY X
(1) Photometric Interpretation : MONOCHROME1 (M1) or MONOCHROME2 (M2)
(2) Formats supported by Musica1 processing
(3) Formats supported by Musica2 processing
(7) WinCenter/WinWidth are not present. Instead LUT Data is present.
(9) Values are in pixel values (eg. 16384, 16384)
(12) After application of the VOI Lut, the bit depth will be equal to the bit depth before application of the VOI Lut, i.c 12 bit
(15) After application of the VOI Lut, the bit depth will be equal to the bit depth before application of the VOI Lut, i.c 15 bit
The above table should be read as follows:
- This table contains the properties for each DICOM CR-Image Type. The relevant DICOM tags
are shown in the column headers and the values that define a specific DICOM CR-Image Type
can be found below the corresponding tag.
- The PACS DICOM CR-Image Types supported by "Workstation" are marked with an “X”.
2.4.1.2.2 RIS
The "Workstation Service & Configuration Tool" must be used to set the AE Title, port-number, host-name
and capabilities of the remote Modality Worklist SCP.
Only one single remote Modality Worklist SCP can be defined at the same time.
2.4.1.2.2.1 Configuration of a RIS
2.4.1.2.2.1.1 Query keys
Query keys are used when the RIS connection is configured for DICOM Modality Worklist. In the
configuration Tool, the user is allowed to enter and/or modify certain query keys. Based on these keys, the
worklist will be populated at the next RIS query.