Table 2.2-7: Association Reject Reasons
Result Response Reason/Diag
The Dicom library refuses the association SCU Device not known
The Dicom library refuses the association
Maximum number of association
processing threads exceeded
The Dicom library refuses the association A799
Dicom Library is not allowed to accept
C-STORE commands from this device
and therefore refuses the association.
The Dicom library refuses the association A0102
SOP Class is not found back in the
configuration
The Dicom library refuses the association A702
The Dicom library is unable to create the
Dicom media file due to the fact that the
disk is full
The Dicom library refuses the association A703
The Dicom library is unable to create the
Dicom media file due to resource
problems other than ‘disk is full’.
The Dicom library refuses the association A703
The Dicom library is unable to acquire
the complete C-STORE request due to
network problems.
The Dicom library refuses the association A701
The Dicom library is unable to acquire
the complete C-STORE request due to
network problems (time-out while
reading data from socket)
2
.
2.2.1.3.1.2 Proposed Presentation Contexts
The DICOM Store AE is capable of proposing the Presentation Contexts shown in the following table:
2
The time-out value is hard-coded in the Dicom library as being 60 seconds.