AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 42 of 92
Document No. 001132, Revision 2.0
Agfa HealthCare NodeID Livelink : 14140190 14 June, 2007
· N-EVENT-REPORT
"Workstation" (SCU) can send the following DIMSE services:
· N-GET
N-EVENT-REPORT is used by the SCP to report execution status changes to "Workstation" (SCU) in an
asynchronous way.
N-EVENT-REPORT has the following arguments
:
Table 2.2-33: Notification Event Information
Event Type Name Event Type ID Attribute Tag
Execution Status Info (2100, 0030)
Print Job ID (2100, 0010)
Film Session Label (2000, 0050)
Pending 1
Printer Name (2110, 0030)
Execution Status Info (2100, 0030)
Print Job ID (2100, 0010)
Film Session Label (2000, 0050)
Printing 2
Printer Name (2110, 0030)
Execution Status Info (2100, 0030)
Print Job ID (2100, 0010)
Film Session Label (2000, 0050)
Done 3
Printer Name (2110, 0030)
Execution Status Info (2100, 0030)
Print Job ID (2100, 0010)
Film Session Label (2000, 0050)
Failure 4
Printer Name (2110, 0030)
Note:
The SCU only releases the Association after the receipt of the event type Done or Failure, if the print
job sop class is supported.
"Workstation" (SCU) returns the confirmation from the N-EVENT-REPORT operation.
N-GET is used to retrieve an instance of the Print Job SOP Class. "Workstation" (SCU) uses the N-GET to
request the SCP to get a Print Job SOP Instance. "Workstation" specifies the UID of the SOP Instance to be
retrieved.
Table 2.2-34: Supported N-GET Attributes for a Print Job
Attribute Name Tag Value
Execution Status (2100, 0020) PENDING, PRINTING, DONE, FAILURE