AGFA HE/001132 Medical Alarms User Manual


 
HE/001132 Page 43 of 92
Document No.001132, Revision 2.0
NodeID Livelink : 14140190
Agfa HealthCare 14 June, 2007
Attribute Name Tag Value
Execution State Info (2100, 0030)
Printer dependent.
Print Priority (2000, 0020) HIGH, MED, LOW
Creation Date (2100, 0040) Date of print job creation.
Creation Time (2100, 0050) Time of print job creation
Printer Name (2110, 0030) User defined name identifying the printer
Originator (2100, 0070) DICOM AE title that issued the print operation
2.2.2.3.1.5 SOP Specific Conformance Presentation LUT SOP Class
(1.2.840.10008.5.1.1.23)
The Presentation LUT Information Object is an abstraction of a Presentation LUT. The objective of the
Presentation LUT is to realize image display tailored for specific modalities, applications, and user
preferences. It is used to prepare image pixel data for display on devices that conform to the Grayscale
Standard Display Function.
The output of the Presentation LUT is Presentation Values (P-Values).
A N-CREATE is issued by "Workstation" (SCU) to create a Presentation LUT SOP Instance. The supported
Presentation LUT attributes are listed below:
Table 2.2-35: Supported Attributes for Presentation LUT
Tag Name Supported Default
(2050,0020) Presentation LUT Shape IDENTITY (Note) IDENTITY
Note:
‘Presentation LUT Sequence’ is not supported.
"Workstation" (SCU) will process the N-CREATE confirmation and response Status codes. The status codes
listed below are recognized:
Table 2.2-36: N-CREATE confirmation and response Status codes
Code Status Meaning
0000 Success Presentation LUT successfully created
B605 Warning Requested Min or Max Density outside
of printer’s operating range. The printer will use its
respective minimum or maximum value instead.