A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
77
AD-1653 Density Determination Kit
Example
Temperature Water density
0°C 0.99984 g/cm
3
10°C 0.99970 g/cm
3
20°C 0.99820 g/cm
3
30°C 0.99565 g/cm
3
AD-1683 DC Static Eliminator
The eliminator is able to dissipate static electricity on a
charged sample.
The eliminator does not use a ventilator and can be used
in precision powder weighting.
AX-USB-9P-EX USB Converter
Adds a COM port to a PC.
Enables bi-directional communication between the PC and the balance when a USB driver is
installed.
Can use serial communication software such as WinCT on a PC without COM ports.
An RS-232C cable is provided to connect the USB converter to the balance.
Weigh in air
Weigh in water - Weigh in the air
x water density = sample density
10.0000 g
10.0000 g 9.5334 g
x 0.9970 g/cm
3
= 21.4 g/cm
3
To USB port
on a PC
RS-232C
on the GH series
balance
USB serial
converter
RS-232C cable provided
(
D-sub 9-
p
in
)