A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
49
Calibration report using an external weight
Key operation
1 Press and hold the CAL key to display Cal out and release the key.
2 Cal 0 is displayed.
3 When updating the calibration mass value, press the RANGE key and proceed to step 4.
When using preset calibration mass value, proceed to step 5.
4 Specify calibration mass value using the following keys.
RANGE key
........
The key to select the blinking figure
RE-ZERO (+)key
..
The key to increase the value of the blinking figure.
MODE (-)key
........
The key to decrease the value of the blinking figure.
PRINT key
.........
The key to store the new value.
5 When pressing the RE-ZERO key, the zero point is measured and the weight value is
displayed for a few seconds.
6 Place the displayed mass on the pan. Press the PRINT key to measure the mass and the
weight value is displayed for a few seconds.
7 If GLP output is used, glp is displayed and the calibration report is output.
8 The balance returns to weighing mode after this calibration.
Setting of "info 1" Setting of "info 2"
AD-8121 printer format General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration weight
Signature
-----------A-&-D
MODEL-----GH-300
S/N-----01234567
ID------LAB-0123
DATE--2004/07/01
TIME----12:34:56
CALIBRATED(EXT.)
CAL.WEIGHT
----+200.0000--g
SIGNATURE
---------------