11.2. GLP Report
Set the following parameters to output the report.
Refer to "15.2. Connection to peripheral equipment" for connection to an AD-8121B.
To print the report, set the "GLP output (info)" parameter to "1" and use MODE 3 of the
AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the STAT.
key of the AD-8121B.
To output the report to a personal computer using the RS-232C interface, set the "GLP
output (info)" parameter to "2".
If the time and date are not correct set the correct time and date in "Clock (Cl adj)" of the
function table.
Calibration report using the internal mass
Key operation
1 Press the CAL key to display Cal in and calibrate the balance automatically.
2 If GLP output is used, glp is displayed and the calibration report is output.
3 The balance returns to weighing mode after this calibration.
Setting of "info 1" Setting of "info 2"
AD-8121 printer format General format
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Signature
===========A=&=D
MODEL=====GH-300
S/N=====01234567
ID======LAB-0123
DATE==2004/07/01
TIME====12:34:56
CALIBRATED(INT.)
SIGNATURE
---------------