A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
17
6.2. Dual Range
Weighing range
GH-252 GH-202
Available minimum display
Precision range 0 g to 101 g 0 g to 51 g 0.01 mg 0.1 mg 1 mg
Standard range 101 g to 250 g 51 g to 220 g 0.1 mg 1 mg
The GH-252 and GH-202 are
equipped with two ranges "precision
range" and "standard range".
When weighing is started by
pressing the ON:OFF key, the
minimum display will be 0.1 mg.
Pressing the RANGE key will
switch these ranges alternately.
When a sample is weighed using the
precision range with a minimum display of
0.01 mg and the weight value exceeds the
precision range value, the minimum display
changes to 0.1 mg of the standard range.
When removing the sample, the minimum
display changes to 0.01 mg of the precision
range automatically.
When a tare weight (container mass value)
exceeds the precision range value, even if
the sample is within the precision range
value, 0.01 mg of the precision range can
not be selected for the minimum display. In
order to select 0.01 mg of the precision
range, remove the tare weight and press
the RE-ZERO key to cancel it.
When a minimum display of 0.1 mg or 1 mg is
selected by the RANGE key, the minimum
display is maintained even if the range is
changed during weighing.
Turn on the balance
Turn on the balance
Minimum display
0.1mg
Minimum display
0.01mg
Minimum display
1mg
Precision Range
Minimum display
0.1mg
Minimum display
1mg
Standard Range
Minimum display
0.01mg
Minimum display
0.1mg
Minimum display
0.01mg
Minimum display
0.01mg
Minimum display
0.1mg
Zero display after
removing the container
Minimum display
0.01mg
Zero display after
placing a container
Minimum display
0.1mg