A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
30
9.2. Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode),
100% reference mass value (percent mode)
The data that is stored in the balance using the data memory function
External calibration weight and target mass value
Function switch settings ("9.1. Permit Or Inhibit")
Liquid density and temperature in the density mode
Note Be sure to calibrate the balance after initialization.
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the
RANGE key, press the ON:OFF key to display p5 .
3 Press the RANGE key to display Clr .
4 Press the PRINT key.
To cancel this operation, press the CAL key.
5 Press the RE-ZERO key to display Clr go .
6 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
Press and hold Press