A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
55
Recalling the memory data
Confirm that the "Data memory (data)" parameter is set to "1".
1 Press and hold the
PRINT key until reCall is displayed,
then release the key.
2 Press the PRINT key to enter the memory recall mode. The
type of data appears in the upper left of the display as shown to
the right. Recall the data in memory using the following keys.
RE-ZERO key
.....
To proceed to the next data set.
MODE key
.........
To go back to the previous data set.
PRINT key
.........
To transmit the current data using the
RS-232C interface.
With RANGE held down, press the CAL key
To delete the current data.
CAL key
.............
To exit the memory recall mode.
3 Press the CAL key to return to the weighing mode.
Transmitting all memory data at one time
Confirm that the "Serial interface (5if)" parameters are set properly.
Refer to "10. Function Table" and "15.2. Connection To Peripheral Equipment".
1 Press and hold the PRINT key until reCall is displayed,
then release the key.
2 Press the RANGE key to display out .
3 Press the PRINT key to display out "n"o with "no" blinking.
5 Press the RE-ZERO key to display out "g"o with "go" blinking.
6 Press the PRINT key to transmit all data using the RS-232C
interface.
7 The balance displays Clear when all data is transmitted.
Press the CAL key to return to the weighing mode.
Deleting all memory data at one time
1 Press and hold the PRINT key until reCall is displayed, then release the key.
2 Press the RANGE key several times to display Clear .
3 Press the PRINT key to display Clr "n"o with "no" blinking.
4 Press the RE-ZERO key to display Clr "g"o with "go" blinking.
5 Press the PRINT key to delete all data
6 The balance displays end and returns to the weighing mode.
Indicators
Weighing data
without time and date
Weighing data with
time and date
Indicators
Weighing data
without time and date
Weighing data with
time and date