A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
21
7. Response Adjustment
This function stabilizes the weight value, reducing the influence on weighing that is caused
by drafts and/or vibration at the place where the balance is installed. This function adjusts by
automatically analyzing the environment or by hand-operation. The function has three stages
as follows :
Indicator Parameter Response Stability
FAST Cond 0
MID. Cond 1
SLOW Cond 2
Fast response, Sensitive value
Slow response, Stable value
7.1. Automatic Response Adjustment
This function automatically updates the response adjustment by analyzing the influence of
the environment using the internal mass.
1 Press and hold the MODE key until RESPONSE is
displayed, then release the key.
2 The balance automatically sets the response characteristic.
Caution Do not allow vibration or drafts to affect the
balance during adjustment.
3 After automatic adjustment, the balance displays
1end1,
returns to the weighing mode and displays the updated
response indicator for about thirty seconds.
Note
If the automatic response adjustment fails, the balance displays CH ng . Check the ambient
conditions such as breeze and vibration, also check the weighing pan. Then, perform the
adjustment again. To return to the weighing mode, press the
CAL key.
If there is matter on the weighing pan, the balance displays CH 0 . Remove them from the
pan. To return to the weighing mode, press the CAL key.
Advise
If the automatic response adjustment is not helpful, try "7.2. Manual Response Adjustment".
Response
indicator
Press and hold
Release
Resul
t