10.2. Details of the Function Table
Class Item and Parameter Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment.
0
Stable range is
±
1 digit
1
#1
5t-b
Stability band width
2
Stable range is
±
3 digits
The stabilization indicator illuminates
when the display fluctuation is within
the range.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
ba5fnc
Environment
Display
p-on
Auto display-ON
1
ON
Turns on the weighing mode display
when the AC adapter is connected.
Cl adj
Clock
Refer to "10.7. Clock and Calendar Function"
The time and date are added to the
output data.
0
Key mode
Accepts the PRINT key only when
the display is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of
ap-p
,
ap-b
and the reference value are met.
prt
Data output mode
3
Stream mode /
Interval memory mode
With
data 0
, outputs data
continuously; with
data 2
, uses
interval memory.
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print difference
2
1000 digits
Difference between reference value
and displayed value
#2
0
Not used
1
Stores weighing data
dout
Data output
data
Data memory
2
Stores calibration data
Related items:
prt,
int
,
d-no
,
5-td
,
info
:
Factory settings.
#1 The unit of minimum display is digit.
Example: If 1 mg display is selected using the RANGE key for the GH-300, 1 mg is one
digit.
#2 Usable minimum display of the balance is one digit.
Example: In gram display, one digit is 0.01 mg for the GH-252 and 0.1 mg for the GH-300.