A&D GH-120/GH-200/GH-300/GH-202/GH-252 Scale User Manual


 
35
Class Item and Parameter Description
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minute
7
5 minute
int
Interval time
8
10 minute
Interval time in the interval memory
mode when using
prt 3, data 1
0
No output
d-no
Data number output
1
Output
Refer to "12. Data Memory"
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time
or date is added to the weighting
data. Refer to "10.7. Clock and
Calendar Function" for details.
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not automatic
feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
Refer to "11. ID Number And GLP
Report "
for details.
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically after data
is output
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
5if
Serial
interface
type
Data format
5
CSV format
Refer to "10.5. Description of Item
"Data Format".
:
Factory settings.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending