Teledyne 100E Blood Glucose Meter User Manual


 
TELEDYNE INSTRUMENTS
Model 100EU Instruction Manual
APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-21
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
APPENDIX A-3: Warnings and Test Functions, Revision F.0B
Table A-2: M100EU Warning Messages, Revision F.0B
Name
1
Message Text Description
WSYSRES SYST EM RESET
Instrument was power-cycled or the CPU
was reset.
WDATAINIT DATA INITIALIZED Data storage was erased.
WCONFIGINIT CONFIG INITIALIZED
Configuration storage was reset to factory
configuration or erased.
WPMT PMT DET WARNING
PMT detector outside of warning limits
specified by DETECTOR_LIMIT variable.
WUVLAMP UV LAMP WARNING
UV lamp reading outside of warning limits
specified by DETECTOR_LIMIT variable.
WSAMPFLOW SAMPLE FLOW WARN
Sample flow outside of warning limits
specified by SAMP_FLOW_SET variable.
WSAMPPRESS SAMPLE PRESS WARN
Sample pressure outside of warning limits
specified by SAMP_PRESS_SET
variable.
WVACPRESS
5
VACUUM PRESS WARN
Vacuum pressure outside of warning limits
specified by VAC_PRESS_SET variable.
WBOXTEMP BOX TEMP WARNING
Chassis temperature outside of warning
limits specified by BOX_SET variable.
WRCELLTEMP RCEL L TEMP WARNING
Reaction cell temperature outside of
warning limits specified by RCELL_SET
variable.
WO2CELLTEMP
10
O2 CELL TEMP WARN
O
2
sensor cell temperature outside of
warning limits specified by O2_CELL_SET
variable.
WIZSTEMP IZS TEMP WARNING
IZS temperature outside of warning limits
specified by IZS_SET variable.
WPMTTEMP PMT TEMP WARNING
PMT temperature outside of warning limits
specified by PMT_SET variable.
WDARKCAL
5
DARK CAL WARNING
Dark offset above limit specified by
DARK_LIMIT variable.
WHVPS HVPS WARNING
High voltage power supply output outside
of warning limits specified by HVPS_SET
variable.
WDYNZERO CANNOT DYN ZERO
Contact closure zero calibration failed
while DYN_ZERO was set to ON.
WDYNSPAN CANNOT DYN SPAN
Contact closure span calibration failed
while DYN_SPAN was set to ON.
WREARBOARD REAR BOARD NOT DET
Rear board was not detected during
power up.
WRELAYBOARD RELAY BOARD WARN
Firmware is unable to communicate with
the relay board.
WFRONTPANEL FRONT PANEL WARN
Firmware is unable to communicate with
the front panel.
WANALOGCAL ANALOG CAL WARNING
The A/D or at least one D/A channel has
not been calibrated.