Teledyne 100E Blood Glucose Meter User Manual


 
TELEDYNE INSTRUMENTS
Model 100EU Instruction Manual
APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-11
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
APPENDIX A-2: Setup Variables For Serial I/O, Revision F.0B
Table A-1: M100EU Setup Variables, Revision F.0B
M100EU Setup Variables for Latest Revision
Setup Variable Numeric
Units
Default
Value
Value
Range
Description
DAS_HOLD_OFF
Minutes 15 0.5–20 Duration of DAS hold off period.
TPC_ENABLE
ON,
OFF
4
OFF, ON
ON enables temperature and
pressure compensation; OFF
disables it.
50
RCELL_SET
ºC
Warnings:
45–55
30–70
Reaction cell temperature set
point and warning limits.
50
IZS_SET
1
ºC
Warnings:
45–55
30–70
IZS temperature set point and
warning limits.
DYN_ZERO
OFF OFF, ON
ON enables contact closure
dynamic zero; OFF disables it.
DYN_SPAN
OFF OFF, ON
ON enables contact closure
dynamic span; OFF disables it.
CONC_PRECISION
1 AUTO,
0,
1,
2,
3,
4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
STABIL_GAS
99
SO2 SO2,
O2
10
,
CO2
11
Selects gas for stability test
measurement.
CLOCK_ADJ
Sec./Day 0 -60–60
Time-of-day clock speed
adjustment.
LANGUAGE_SELECT
ENGL ENGL,
SECD,
EXTN
Selects the language to use for
the user interface. Enclose value
in double quotes (“) when setting
from the RS-232 interface.
MAINT_TIMEOUT
Hours 2 0.1–100
Time until automatically
switching out of software-
controlled maintenance mode.
CONV_TIME
33 MS 33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT and UV
detector channels. Enclose value
in double quotes (“) when setting
from the RS-232 interface.
DWELL_TIME
Seconds 1, 0.1–10
Dwell time before taking each