M100EU – OPERATIONS MANUAL TELEDYNE INSTRUMENTS
Addendum to M100E Manual - P/N 04145 CALIBRATION PROCEDURES
7. CALIBRATION PROCEDURES
Calibration of the M100EU should be performed according to the procedures described in Chapters 7 & 8 of the
M100E Manual - P/N 04145. However, delivering span and zero gases for the lower ranges that the M100EU is
designed for can be difficult. For best results when calibrating the M100EU, wait one hour for the instrument to
stabilize when delivering zero and span gases before pressing the zero and span buttons. Attention must be
paid to the quality of the gasses, the level of contaminants in the gases as well as the history and conditioning of
the gas delivery components. Only Teflon or glass should be used for any “wetted” surfaces that the calibration
gasses contact. All delivery system components should be conditioned by running span gas for a minimum of
four hours before conducting actual span calibrations.