O2_FILT_DELTA
10
% 2 0.1–100
Absolute change in O
2
concentration to shorten filter.
O2_FILT_PCT
10
% 2 0.1–100
Relative change in O
2
concentration to shorten filter.
O2_FILT_DELAY
10
Seconds 20 0–300
Delay before leaving O
2
adaptive
filter mode.
O2_DIL_FACTOR
10
1 0.1–1000
Dilution factor for O
2
. Used only if
is dilution enabled with
FACTORY_OPT variable.
50
O2_CELL_SET
10
ºC
Warnings:
45–55
30–70
O
2
sensor cell temperature set
point and warning limits.
O2_STD_CELL_TEMP
10
ºK 323 1–500
Standard O
2
cell temperature for
temperature compensation.
O2_STD_CELL_PRESS
10
"Hg 28.50 1.00–50.00
Standard O
2
cell pressure for
pressure compensation.
PPB,
PPM,
UGM,
MGM
USER_UNITS
PPB,
PPM
3
PPM,
MGM
3
Concentration units for user
interface. Enclose value in
double quotes (“) when setting
from the RS-232 interface.
LAMP_DRIVE
6
mV 5000 0–5000 Lamp power setting.
LAMP_CAL
mV 3500 1000–5000 Last calibrated UV lamp reading.
LAMP_GAIN
0.9 0.5–1.5
UV lamp compensation
attenuation factor.
BXTEMP_TPC_GAIN
1,
0
12
0–10
Box temperature compensation
attenuation factor.
SPRESS_TPC_GAIN
1 0–10
Sample pressure compensation
attenuation factor.
PMT_TARG_CONC
12
Conc 400 0.01–9999.99
Target SO
2
concentration during
PMT calibration.
PMT_UPDATE_PERIOD
12
Seconds 10 1–100
Period between HVPS gain
updates during PMT calibration.
PMT_CAL_TIMEOUT
12
Minutes 5 1–100
Maximum time for PMT
calibration to succeed.
HVPS_ADJUST
12
0 0–200 HVPS gain adjustment.
HVPS_INTEG
12
Gain 5 0–500
Integral coefficient for adjusting
HVPS gain during PMT
calibration.
HVPS_STABIL
12
1 0.1–10
HVPS gain must stabilize to
within this limit for PMT
calibration to succeed.
PMT_ADJUST
12
0 0–65535 PMT gain adjustment.
SLOPE_CONST
8,
6.25
3
0.1–10
Constant to make visible slope
close to 1.