Teledyne 100E Blood Glucose Meter User Manual


 
TELEDYNE INSTRUMENTS
Model 100EU Instruction Manual
APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-19
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
DIM
I2C_RESET_ENABLE
ON OFF, ON I
2
C bus automatic reset enable.
CLOCK_FORMAT
“TIME=%H:%
M:%S”
Any character
in the allowed
character set.
Up to 100
characters
long.
Time-of-day clock format flags.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
“%a” = Abbreviated weekday
name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format
(00 – 23).
“%I” = Hour in 12-hour format (01
– 12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal
number (01 – 12).
“%M” = Minute as decimal
number (00 – 59).
“%p” = A.M./P.M. indicator for
12-hour clock.
“%S” = Second as decimal
number (00 – 59).
“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as
decimal number.
“%%” = Percent sign.
FACTORY_OPT
BitFlag 0 0–65535
Factory option flags. Add values
to combine flags.
1 = enable dilution factor
2 = zero/span valves installed
4 = IZS installed (implies
zero/span valves installed)
8 = low span valve installed
16 = display units in
concentration field
32 = enable software-controlled
maintenance mode
64 = enable lamp power analog
output
128 = enable switch-controlled
maintenance mode
256 = compute only offset during
zero calibration
1024 = enable high flow rate
sensor