APPENDIX A - Version Specific Software Documentation
05928 Rev F.0B A-17
DCN 5063 PRINTED DOCUMENTS ARE UNCONTROLLED
19200,
38400,
57600,
115200
MODEM_INIT2
“AT Y0 &D0
&H0 &I0 S0=2
&B0 &N6 &M0
E0 Q1 &W0”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 COM2 modem
initialization string. Sent verbatim
plus carriage return to modem on
power up or manually. Enclose
value in double quotes (“) when
setting from the RS-232
interface.
RS232_PASS
Password 940331 0–999999 RS-232 log on password.
MACHINE_ID
ID 100 0–9999 Unique ID number for instrument.
COMMAND_PROMPT
“Cmd> ”
Any character
in the allowed
character set.
Up to 100
characters
long.
RS-232 interface command
prompt. Displayed only if enabled
with RS232_MODE variable.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
TEST_CHAN_ID
NONE NONE,
PMT
READING,
UV READING,
VACUUM
PRESSURE
3
,
SAMPLE
PRESSURE,
SAMPLE
FLOW,
RCELL TEMP,
O2 CELL
TEMP
10
,
CHASSIS
TEMP,
IZS TEMP
1
,
PMT TEMP,
HVPS
VOLTAGE
Diagnostic analog output ID.
Enclose value in double quotes
(“) when setting from the RS-232
interface.
REMOTE_CAL_MODE
LOW LOW,
HIGH,
CO2
11
,
O2
10
Range to calibrate during
contact-closure and Hessen
calibration. Enclose value in
double quotes (“) when setting
from the RS-232 interface.
PASS_ENABLE
OFF OFF, ON
ON enables passwords; OFF
disables them.
STABIL_FREQ
99
Seconds 10 1–300
Stability measurement sampling
period.
STABIL_SAMPLES
99
Samples 25 2–40
Number of samples in
concentration stability reading.
RCELL_CYCLE
Seconds 2 0.5–30
Reaction cell temperature control
cycle period.