Model 100EU Instruction Manual APPENDIX A - Version Specific Software Documentation
A-18 05928 Rev F.0B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5063
RCELL_PROP
1/ºC
0.3 (prop. band
= 3.3 ºC)
0–10
Reaction cell temperature PID
proportional coefficient.
RCELL_INTEG
0.005 0–10
Reaction cell temperature PID
integral coefficient.
RCELL_DERIV
0.5 0–10
Reaction cell temperature PID
derivative coefficient.
O2_CELL_CYCLE
10
Seconds 10 0.5–30
O
2
cell temperature control cycle
period.
O2_CELL_PROP
10
1 0–10
O
2
cell PID temperature control
proportional coefficient.
O2_CELL_INTEG
10
0.1 0–10
O
2
cell PID temperature control
integral coefficient.
O2_CELL_DERIV
10
0 (disabled) 0–10
O
2
cell PID temperature control
derivative coefficient.
IZS_CYCLE
1
Seconds 2 0.5–30
IZS temperature control cycle
period.
IZS_PROP
1
1/ºC
1 (prop. band
= 1 ºC)
0–10
IZS temperature PID proportional
coefficient.
IZS_INTEG
1
0.03 0–10
IZS temperature PID integral
coefficient.
IZS_DERIV
1
0 0–10
IZS temperature PID derivative
coefficient.
650,
550
3
HVPS_SET
Volts
Warnings:
400–900,
400–700
3
0–2000
High voltage power supply
warning limits. Set point is not
used.
MAX_PMT_DETECTOR
mV 499 5 0–5000
PMT detector maximum warning
limit.
450
PHOTO_ABS_LIMITS
1
mV
Warnings:
125–625
0–5000
Pre-amplified UV lamp
minimum/maximum warning
limits. Set point is not used.
3500
UV_LAMP_LIMITS
mV
Warnings:
1000–4995
0–5000
UV lamp minimum/maximum
warning limits. Set point is not
used.
ELEC_TEST_LEVEL
12
0 0–65535 Electrical test level setting.
OPTIC_TEST_LEVEL
12
0 0–65535 Optical test level setting.
CONC_LIN_ENABLE
3
ON OFF, ON
ON enables concentration
linearization; OFF disables it.
SERIAL_NUMBER
“000 00000
Any character
in the allowed
character set.
Up to 100
characters
long.
Unique serial number for
instrument. Enclose value in
double quotes (“) when setting
from the RS-232 interface.
DISP_INTENSITY
HIGH HIGH,
MED,
LOW,
Front panel display intensity.
Enclose value in double quotes
(“) when setting from the RS-232
interface.