AUTO2
PHYS_RANGE1
PPM 2,
500
3
0.1–2500,
5–10000
3
Low pre-amp range.
PHYS_RANGE2
PPM 22,
5500
3
0.1–2500,
5–10000
3
High pre-amp range.
CONC_RANGE1
Conc 500,
5000
3
0.1–50000 D/A concentration range 1.
CONC_RANGE2
Conc 500,
5000
3
0.1–50000 D/A concentration range 2.
CO2_RANGE
11
% 15 0.1–500 CO
2
concentration range.
O2_RANGE
10
% 100 0.1–500 O
2
concentration range.
700,
250
1+9
SAMP_FLOW_SET
cc/m
Warnings:
350–1200,
175–325
1+9
0–6000
Sample flow set point for flow
calculation and warning limits.
SAMP_FLOW_SLOPE
1 0.5–1.5
Sample flow slope correction
factor (adjusted flow = measured
flow x slope).
VAC_SAMP_RATIO
3
0.53 0.1–2
Maximum vacuum pressure /
sample pressure ratio for valid
sample flow calculation.
29.92
SAMP_PRESS_SET
"Hg
Warnings:
15–35
0–100
Sample pressure set point for
pressure compensation and
warning limits.
6
VAC_PRESS_SET
3
"Hg
Warnings:
3–10
0–100
Vacuum pressure set point for
pressure compensation and
warning limits.
30
BOX_SET
ºC
Warnings:
8–50
5–60
Box temperature warning limits.
Set point is not used.
7,
15
12
PMT_SET
ºC
Warnings:
2–12,
2–20
12
0–40
PMT temperature set point and
warning limits.