Sanyo 550 Thermometer User Manual


 
Description
Type
Source
Header
Data Block
SET CONTROLLER TO REMOTE ACCESS MODE
Command
master
136
No
This instructs the Controller to ignore key presses on its local keyboard and
display "Remote" in the Access window.
Description
Type
Source
Header
Data Block
SET CONTROLLER TO REMOTE ACCESS MODE
Reply
Controller
136
No
This packet is sent by the Controller to confirm to the master that the Controller is
switched to Remote Access Mode.
Description
Type
Source
Header
Data Block
SET CONTROLLER TO LOCAL ACCESS MODE
Command
master
137
No
This instructs the Controller to allow key presses on its local keyboard and dis
play "Local" in the Access window.
Description
Type
Source
Header
Data Block
SET CONTROLLER TO LOCAL ACCESS MODE
Reply
Controller
137
No
This packet is sent by the Controller to confirm to the master that the Controller is
switched to Local Access Mode.
Description
Type
Source
Header
Data Block
SET/CHANGE STEADY STATE CONDITION
Command
master
138
20 bytes
This command instructs the Controller to run a steady state set point. The data
block parameters are as follows:
- 6 bytes: Temperature Set Point in °C, eg"-150.0". There should be no spaces in
this string; zeros should be used to pad the string, e.g. -10°C should be
represented as "-010.0". 4.3°C should be represented as "0004.3".
- 3 bytes: Relative Humidity Set Point in %, eg "100". There should be no spaces
in this string; e.g. 0%RH should be represented as "000".
- 8 bytes: Digital Channels, eg "10010011".
60