4.INSTALLATION
4.1.
ACCESSORIES & SPARES
Check that the following items have been supplied with the cabinet.
a) Wick for the wet bulb sensor(s)
b) A set of psychrometric tables for the recorder (if fitted)
c) Instructions and wick for recorder (if fitted)
4.2.
POSITIONING
4.2.1 Cabinets not requiring on-site commissioning
Place the cabinet on a level site, or suitable surface in the case of bench standing models,
in such a position that air is free to circulate around it. The cabinet should be placed at
least 450mm (18ins) away from any obstruction, eg: a wall or workbench.
For floor standing models lock the front castors to prevent movement.
Stand the cabinet in its operating position for 24 hours before switching on, to allow oil
in the refrigeration system to return to sump. If the cabinet is moved, allow it to stand
again.
If required, we can arrange for one of our engineers to commission the cabinet. Should
you feel that his services are required please telephone us at Loughborough to make the
necessary arrangements. During his visit our engineer will run the cabinet and check
that its systems are functioning correctly and enable you assure that your staff are fully
acquainted with the cabinet and its operation.
4.2.2 Cabinets requiring on-site commissioning
Cabinets supplied with a separate or remote refrigeration unit should be positioned to
enable the pipe and cable connections to be made by our service engineer. Positioning
must be on a level site such that the cabinet and refrigeration unit are at least 450mm
(18 ins) away from any obstruction, eg: a wall or workbench. Lock the front castors to
prevent movement.
During his visit our engineer will run the cabinet and check that its systems are
functioning correctly. Take advantage of the commissioning visit to ensure that your
staff are fully acquainted with the cabinet and its operation.
Connect a suitable power supply to the cabinet (see section 4.4). The maximum power
ratings are shown on the cabinet rating plate (and refrigeration unit rating plate, if
applicable).
DO NOT SWITCH ON THE CABINET UNTIL IT HAS BEEN
COMMISSIONED.
12